Attest Super Rapid Readout Biological Indicators, 50/Pkg 1492V
Attest Super Rapid Readout Biological Indicators, 50/Pkg 1492V - 50 Vials/Box is backordered and will ship as soon as it is back in stock.
Couldn't load pickup availability
Delivery and Shipping
Delivery and Shipping
Items in stock will ship out for 2-3 business day delivery time. Items not in stock can take up to but not limited to 4 weeks, please contact us at 1-855-512-9159 to determine if an item is in stock
Description
Description
Attestā¢ Super Rapid Biological Indicators monitor the efficacy of steam sterilization processes with quick and reliable results.Ā The 3Mā¢ Attestā¢ Super Rapid Readout Biological Indicator 1492V (brown cap, referred to hereinafter as the 1492VĀ BI) is a self-contained biological indicator specifically designed for rapid and reliable qualification testing and routine monitoring of 270Ā°F (132Ā°C) and 275Ā°F (135Ā°C) dynamic-air-removal (pre-vacuum) steam sterilization processes when used in conjunction with the 3Mā¢ Attestā¢ Auto-readerĀ 490 (hereinafter referred to as the 490 Auto-reader). The 1492VĀ BI is a single-useĀ device.
Ā
Readout Times:Ā
The 1-hour super rapid readout and the optional 48-hour visual pH color change incubation times have been correlated with a 7-day incubation period (at 56+/-2Ā°C) following the FDAās Reduced Incubation Time protocol. Processed indicators were examined at 48 hours and 7 days for detection of a visual pH color change.
The 1-hour fluorescence change readings and the 48-hour visual pH color change readings were compared to the 7-day visual pH color change readings to determine the readout time of theĀ indicator. 1-hour Fluorescence ChangeĀ Result 1492VĀ BIs have 1-hour reduced incubation time results that correlate to the 7-day (168 hours) visual readout result ā„ 97% of theĀ time. 48-hour Visual pH Color ChangeĀ Result 1492VĀ BIs have 48-hour reduced incubation time results that correlate to the 7-day (168 hours) visual readout result ā„ 97% of theĀ time. Due to the high reliability of the 1-hour fluorescent result there is no advantage to incubating 1492VĀ BIs beyond 1Ā hour. 1492VĀ BIs meet ANSI/AAMI/ISO 11138-1:2006/(R)2010, ANSI/AAMI/ISO 11138-3:2006/(R)2010 and BS EN/ISO 11138-1:2006, BS EN/ISOĀ 11138-3:2006.
Ā
- Provides BI results in 24 minutes in 3M Attest 490/490H Auto-readers with software version 4.2.7 or greater
- Dynamic-air-removal (Pre-vacuum and SFPP): 132ĖC (270ĖF) 3 or 4 min. 135ĖC (275ĖF) 3 min.
- Meets ISO and FDA performance requirements for biological indicators
- Contains viable organisms that provide a direct measure of lethality
- Puts your instrument turnaround speed into overdrive
- Puts your instrument turnaround speed into overdrive for improved patient safety
- *When used with 3Mā¢ Attestā¢ Auto-reader 490 or the Attestā¢ Auto-reader 490H with software version 4.2.7 or greater.
-
Twenty-four minute readout Biological Indicator for 132Ā°C (270Ā°F) 3 or 4 minute and 135Ā°C (275Ā°F) 3 minute Dynamic-Air-Removal Steam Sterilization cycles (Brown Cap). 50 each/box.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.